import time
class hmp4040():
def __init__(self, pyvisa_instr):
self.hmp4040 = pyvisa_instr # this is the pyvisa instrument, rm.open_resource('ASRL6::INSTR')
self.channel_list = [1, 2, 3, 4]
def get_inst_state(self):
all_scpi_list = []
for channel in self.channel_list:
self.hmp4040.write('INSTrument:NSELect {0}'.format(channel))
voltage = float(self.hmp4040.query('SOURce:VOLTage?'))
current_lim = float(self.hmp4040.query('SOURce:CURRent?'))
status = int(self.hmp4040.query('OUTPut:STATe?'))
current = float(self.hmp4040.query('MEASure:CURRent?'))
volt_prot = float(self.hmp4040.query('VOLTage:PROTection?'))
volt_prot_active = self.hmp4040.query('VOLTage:PROTection:MODE?').rstrip('\r\n')
print "Channel %d voltage is %.02f V, overvoltage limit is %.02f overvolt config is %s, current limit is %.02f A, current is %.03f A status is %s" % (channel,voltage,volt_prot,volt_prot_active,current_lim,current,status)
def get_channel_scpi_list(self, channel=1):
result_list = []
self.hmp4040.write('INSTrument:NSELect {0}'.format(channel))
result_list.append('INSTrument:NSELect {0}'.format(channel))
for command in self.cmd_dict:
result = (self.hmp4040.query(command.format("?"))).rstrip('\r\n')
result = " " + result
result_list.append(command.format(result))
time.sleep(0.1)
return result_list
def get_unique_scpi_list(self):
unique_scpi_list = []
unique_scpi_channel_list = []
for channel in self.channel_list:
channel_settings_list = self.get_channel_scpi_list(channel)
found = 0
unique_scpi_channel_list = []
for setting in channel_settings_list:
if (setting not in self.por_scpi_list):
found = 1
unique_scpi_channel_list.append(setting)
if (found == 1):
unique_scpi_list.append(channel_settings_list[0]) # this is the channel selection
unique_scpi_list.extend(unique_scpi_channel_list)
return unique_scpi_list
select_cmd_dict = {
"INSTrument:NSELect {0}": "Selects a channel by number" }
activate_cmd_dict = {
"OUTPut:GENeral {0}": "Enables the outputs for all activated channels",
"OUTPut:SELect {0}": "Activites a selected channel" }
cmd_dict = {
"SOURce:VOLTage{0}" : "Sets/Queries the voltage value of the selected channel",
"SOURce:CURRent{0}" : "Sets/Queries the current limit value of the selected channel",
"VOLTage:PROTection:MODE{0}" : "Sets/Queries the voltage protection mode, measured or protected",
"OUTPut:STATe{0}" : "Activates and enables the output for a selected channel"
}
query_cmd_dict = {
"MEASure:VOLTage?" : "Measures voltage of selected channel",
"MEASure:CURRent?" : "Measures voltage of selected channel"
}
por_scpi_list = [
'INSTrument:NSELect 1',
'INSTrument:NSELect 2',
'INSTrument:NSELect 3',
'INSTrument:NSELect 4',
'SOURce:CURRent 0.1000',
'SOURce:VOLTage 0.000',
'VOLTage:PROTection:MODE measured',
'OUTPut:STATe 0' ]