Alarm Information

kubu4 edited this page Feb 3, 2021 · 2 revisions

These pages no longer updated, as of 20210126.


Description of active alarms in the lab, how to configure, and what phone calls mean.

#Alarm: Sensaphone Model 1104 User Manual here

Incoming Phone Number: 1-206-685-7806

This allows for you to call the alarm and receive updates on connected alarms, as well as listen to what is happening inside of the room for 15 seconds.

Current numbers programmed to be called in the event of an alarm state:

  • 1: 206-866-5141 : Steven
  • 2: Sam
  • 3: 206-914-3735 :
  • 4: 206-685-3273 : Sam

Alarms:

  • 1: Temperature sensor for ambient temperature inside of room.
  • 2: Power on/off sensor for power being supplied to freezer 1
  • 3: Power on/off sensor for power being supplied to freezer 2
  • 4: Not hooked to anything.

Alarm meaning and acknowledgement:

The alarm indicates only lack of power being supplied to freezer 1. To acknowledge receipt of the alarm, press "555" at any >time during the recorded message. This only acknowledges receipt, the alarm condition still exists.

Alarm sequence:

Currently, When the power goes off, Alarm 2 waits 10 minutes, then issues an alarm status. It then waits 5 further minutes >to begin calling numbers on the list. There is a 1 minute pause between phone number calls. Upshot: We currently do not >know of a power outage until a minimum of 15 minutes after it happens.

To add/change phone numbers (The sensaphone can hold 4 phone numbers):

  • Press Set
  • Press Phone Number
  • Press Number on number pad corresponding to number you would like to change
  • Enter Phone number on number pad
  • Press Enter
  • Sensaphone should respond with "Enter" if number updated correctly.

VWR Fridge Alarm set points: -60* High, -90* Low.


#Alarm: Avtech

Alarms:

  • 1: FTR-209 Freezer [High Alarm: 5, Low Alarm: -50]
  • 2: FTR-209 Refrigerator [High Alarm: 15, Low Alarm: -5]
  • 3: FTR-213 Freezer [High Alarm: 5, Low Alarm: -50]
  • 4: FTR-213 Refrigerator [High Alarm: 15, Low Alarm: -5]

Current numbers programmed to be called in the event of an alarm state:

  • 1: 206-866-5141 : Steven
  • 2: 206-685-3273 : Sam