package "SerialInterfacer"
version "v1.1, Nov. 2018 [Jack J Miller, University of Oxford]"
purpose "This dodgy command line tool acts as a command line interfacer with the Bruker BVT3000 NMR heater unit that uses a Eurotherm 902s PID controller.
It provides many silly, but hopefully useful, command line commands to check hardware status, measure things, and so on. Everything is either dreadfully verbose (with the -v switch), or alternatively consists of:
'***XXXX: VALUE', where 'XXXX' depends on the command requested (e.g. TEMP) and 'VALUE' is the return of that command. See the documentation (that I haven't completely yet written...) for more details. Further documentation is available with the --full-help option.\n
"
text "\n Written by Jack J. Miller, University of Oxford, heavily ''inspired by'' code from Fsc2, a free spectrometer driving software, written by Jens Thoms Toerring. \n
Licensed under under the terms of the GNU General Public License, v3, or at your choice any later license.\n
Given that this software can be used with hardware designed to cool samples to 77 K or heat them to 1000 K, please particularly note the section of the GPL disclaiming liability!\n"
usage "-d /path/to/serialPort command [command argument]"
description "For variable temperature NMR experiments."
option "verbose" v "Debugging verbosity" flag off
#Boring options
section "Serial devices"
option "device" d "Serial port device to use" string required default="/dev/null"
option "list-devices" - "List found serial devices for debugging purposes (specify a dummy -d=Path)" flag off
#Heater
section "Heater controls"
option "heater-on" - "Turn on the heater (Be careful!)" flag off
option "heater-off" O "Turn the heater off" flag off
text "\n"
option "get-heater-state" G "See if the heater is on or off" flag off
option "set-heater-power-limit" - "Live dangerously by increasing the maximum heater power (a percentage between 0/100)" optional float hidden
option "get-heater-power-limit" - "Get the heater power limit" flag off hidden
option "get-heater-power" - "Get the current heater power (as a percentage)" flag off
option "set-heater-power" - "Set the current heater power (as a percentage)" optional float
option "check-heater" X "Check to see if the heater is overheating" flag off
#Gas flow
section "Gas flow controls"
option "get-gas-flow-rate" g "Return the gas flow rate in l/hr" flag off
option "set-gas-flow-rate" s "Set the gas flow rate to one of 16 values between 0 and 2000 l/hr. NB: 0 disables the heater." optional float
#Temperature
section "Temperature methods"
option "read-temperature" r "Read temperature" flag off
option "set-temperature-setpoint" - "Set the temperature setpoint" optional float
option "get-temperature-setpoint" R "Get the current temperature setpoint" flag off
#LN2
section "Liquid Nitrogen methods (NB: you need the optional N2 evaporator)"
option "get-ln2-heater-state" - "Get the LN2 heater state (on/off)" flag off hidden
option "set-ln2-heater-state" - "Turn on/off the LN2 heater (1=on)" optional int hidden
option "get-ln2-heater-power" - "Get the LN2 heater power" flag off hidden
option "set-ln2-heater-power" - "Set the LN2 heater power (as a percentage)" optional float hidden
option "check-ln2-heater" - "Check the LN2 tank to see if it is okay, needs a refil, or is empty" flag off hidden
section "PID control"
option "enable-PID-control" E "Enable the PID control of temperature to the setpoint" flag off hidden
option "manual-mode" M "Set the device into manual heater power mode" flag off hidden
option "get-mode" - "Return the current PID / Manual control mode\n" flag off hidden
option "set-proportional-band" - "Set the P part of PID" optional float hidden
option "get-proportional-band" - "Get the P part of PID" flag off hidden
option "set-integral-time" - "Set the I part of PID" optional float hidden
option "get-integral-time" - "Get the I part of PID" flag off hidden
option "set-differential-time" - "Set the D part of PID" optional float hidden
option "get-differential-time" - "Get the D in the PID\n" flag off hidden
option "get-high-cutback" - "Get the high cutback value" flag off hidden
option "set-high-cutback" - "Set the high cutback value" optional float hidden
option "get-low-cutback" - "Get the low cutback value" flag off hidden
option "set-low-cutback" - "Set the low cutback value" optional float hidden
option "get-adaptive-tune-level" - "Get the adaptive tune level (K)" flag off hidden
option "set-adaptive-tune-level" - "Set the adaptive tune level (K)" optional float hidden
#Misc
section "Misc. methods"
option "lock-keypad" - "Unlock (1) or lock (0) the Eurotherm keypad" optional int hidden
option "get-eurotherm-status" - "Get the status of the Eurotherm controller (alarming or not)" flag off
option "status-all" - "Return the status of everything that is a status" flag off
option "check-sensor-break" - "Check to see if the Thermocouples are broken" flag off
text "\n Example invocation to read temperature (K), and gas flow rate (l/hours): \n\n BVTserialInterfacer -d /dev/ttyUSB0 -r --get-gas-flow-rate\n"