FDA Preparation

Your Name: Elliot Y.

Name of your Device: Machine Learning Model for Pneumonia Detection on Chest X-Rays

Algorithm Description

1. General Information

Intended Use Statement:
This algorithm is comprised of a Convolutional Neural Network model to analyze Chest X-Ray images to detect the presence of pneumonia. It is intended to assist radiologists in the clinical setting and should be used in conjunction with a Radiologist's review of the Xray, sputnum cultures, and a review of a patient's medical history for diagnostic validation.

Indications for Use:
This algorithm may be applied to patients who's medical history may indicate pneumonia infection, of ages 1 thru 95 and of gender Male or Female. X-Ray images must be taken with a PA or AP viewing position.

Device Limitations:
This algorithm must be run with a computer that meet minimum GPU and RAM requirements.
Furthermore, presence of Infiltration may confound device performance. Infiltration may cause the device to classify an X-Ray image as Positive Pneumonia.

Clinical Impact of Performance:
The goal of this device is assist Radiologists with their review of X-Rays for indications of Pneumonia. A Radiologist is required to review all X-Rays and validate results.
When this device incorrectly predicts a False Positive, a patient may be directed for a sputnum culture and the diagnosis can be validated with the sputnum culture.
On the other hand, a False Negative can adversely impact a patient. If the False Negative is not revised by a Radiologist, a clinician may proceed with an incorrect diagnosis and not proceed to sputnum cultures.
Therefore, it is critical for Radiologists to review and validate all results.

2. Algorithm Design and Function

DICOM Checking Steps:
Check DICOM Headers for:
Modality == 'DX'
BodyPartExamined=='CHEST'
PatientPosition in 'PA' or'AP' Position
If any of these three categories do not match their respective requirements, then a message will state that the DICOM does not meet criteria.

Preprocessing Steps:
For DICOMs that pass the intial Header check, the DICOM pixel array will be opened. A copy of the DICOM pixel array will be normalized and then resized to 224 by 224 pixels.

CNN Architecture:

A Sequential Model was built by Fine-tuning the VGG16 Model with ImageNet weights.
This model takes the VGG16 model layers up to and including the block5_pool.
Layers taken from the VGG16 model were frozen, so that their weights were not trained. Output from this pre-trained model was flattened.
These following layers were added:
(Dense(1024, activation = 'relu'))
(Dropout(0.5))
(Dense(512, activation = 'relu'))
(Dropout(0.5))
(Dense(256, activation = 'relu'))
(Dropout(0.5))
(Dense(1, activation = 'sigmoid'))
Output

Full Model architecture:

Layer (type) Output Shape Parameter


input_5 (InputLayer) (None, 224, 224, 3) 0
block1_conv1 (Conv2D) (None, 224, 224, 64) 1792
block1_conv2 (Conv2D) (None, 224, 224, 64) 36928
block1_pool (MaxPooling2D) (None, 112, 112, 64) 0
block2_conv1 (Conv2D) (None, 112, 112, 128) 73856
block2_conv2 (Conv2D) (None, 112, 112, 128) 147584
block2_pool (MaxPooling2D) (None, 56, 56, 128) 0
block3_conv1 (Conv2D) (None, 56, 56, 256) 295168
block3_conv2 (Conv2D) (None, 56, 56, 256) 590080
block3_conv3 (Conv2D) (None, 56, 56, 256) 590080
block3_pool (MaxPooling2D) (None, 28, 28, 256) 0
block4_conv1 (Conv2D) (None, 28, 28, 512) 1180160
block4_conv2 (Conv2D) (None, 28, 28, 512) 2359808
block4_conv3 (Conv2D) (None, 28, 28, 512) 2359808
block4_pool (MaxPooling2D) (None, 14, 14, 512) 0
block5_conv1 (Conv2D) (None, 14, 14, 512) 2359808
block5_conv2 (Conv2D) (None, 14, 14, 512) 2359808
block5_conv3 (Conv2D) (None, 14, 14, 512) 2359808
block5_pool (MaxPooling2D) (None, 7, 7, 512) 0
flatten_4 (Flatten) (None, 25088) 0
dense_15 (Dense) (None, 1024) 25691136
dropout_12 (Dropout) (None, 1024) 0
dense_16 (Dense) (None, 512) 524800
dropout_13 (Dropout) (None, 512) 0
dense_17 (Dense) (None, 256) 131328
dropout_14 (Dropout) (None, 256) 0
dense_18 (Dense) (None, 1) 257

3. Algorithm Training

Parameters:

Keras.preprocessing.image ImageDataGenerator was used with the following parameters to augment training images. my_train_idg = ImageDataGenerator(rescale = 1./255, horizontal_flip = True, height_shift_range = 0.1, width_shift_range = 0.1, rotation_range = 15, shear_range = 0.1, zoom_range = 0.1, samplewise_center = True, samplewise_std_normalization = True )

The Image Data Generator was applied to the training data. The training data was resized to 224x224 and divided into batch size of 16.

train_gen = my_train_idg.flow_from_dataframe(dataframe=train_df, directory=None, x_col = 'path', y_col = 'pneumonia_class', class_mode = 'binary', target_size = (224,224), batch_size = 16 )

The optimizer Adam was applied to the training dataset with a learning rate of 0.001 for two models and 0.0001 for a third model. Binary Cross Entropy was the loss parameter.

optimizer = Adam(learning_rate = 1e-3) loss = 'binary_crossentropy' metrics = ['binary_accuracy']

Model 3 Training Performance Visualizations
Model 3 Training Performance
Figure 1. Model Training History

ROC Curve
Figure 2. ROC Curve for threshold selection by maximizing distance away from chance prediction

F1Score vs. Thresholds
Figure 3. F1-Scores vs Thresholds for threshold selection by maximizing F1-Score

P-R Curve
Figure 4. Precision-Recall Curve for threshold selection by maximizing Recall

Confusion Matrix: Optimize by ROC determined Threshold:

ROC Thresh Positive Negative
Positive 219 701
Negative 67 443

Optimize by F1 determined Threshold:

F1 Thresh Positive Negative
Positive 218 692
Negative 68 452

Maximize Recall by Selecting Threshold with 0.80 Recall, Threshold of 0.355:

ROC Thresh Positive Negative
Positive 124 351
Negative 162 793

Final Threshold and Explanation:
The performance of three thresholds was explored by optimizing by ROC, by F1, and by maximizing Recall.
1. When optimizing using a threshold determined by optimizing ROC (0.24046), its confusion matrix shows it yields 219 TP, 67 FN, 701 FP.
2. When optimizing using a threshold determined by optimizing F1 (0.24509), its confusion matrix shows it yields 218 TP and 68 FN, 692 FP. The performance is similar to optimizing by ROC. With this threshold, the F1 score is 0.36.
3. Based on a Recall of 0.80, the threshold value of 0.355 was chosen. It's corresponding confusion matrix, it yields 124 TP, 162 FN, with 351 FP.
For this project, the first model is the best architecture and its optimal threshold value is 0.24509 as determined from F1. This combination yields a F1 score of 0.36.

4. Databases

Based on EDA work, the Data_Entr_2017.cvs contains 1,431 Images positive for Pneumonia and 110,689 Images negative for Pneumonia.

The data set was split into 80% Training data and 20% for validation data.

Age
Figure 5. Age Distribution of patients with pneumonia in the complete dataset

Description of Training Dataset:
For the training data set, the positive to negative images must be equal in number.
Training DataSet: Pneumonia Positive (80% of all Pneumonia Positive cases in dataset) = 1145 counts, Pneumonia Negative = 1145 counts
Demographics such as age and gender reflect the general data set.

The general dataset showed that the age distribution of Males are statistically different from the age distribution of Females.
A Two-sample T-Test showed that the age distribution of Males in the Training DataSet match the age distrubition of Males in the General Dataset.
Another Two-sample T-Test showed that the age distribution of Females in the Training DataSet match the age distrubition of Females in the General Dataset.

Description of Validation Dataset:
For the validation data set, the demographics must reflect the general data set.
Furthermore, in the clinical setting where this algorithm will be deployed, patients are being X-Rayed based on their clinical symptoms that make Pneumonia highly likely. The prevalence of Pneumonia is about 20% of those who are X-Rayed.

Validation DataSet: Pneumonia Positive (20% of all Pneumonia Positive cases in dataset, no overlap with Training Dataset) =286 counts, Pneumonia Negative = 4*286=1,144 counts
A Two-sample T-Test showed that the age distribution of Males in the Training DataSet match the age distribution of Males in the General Dataset. Another Two-sample T-Test showed that the age distribution of Females in the Training DataSet matches the age distribution of Females in the General Dataset.

5. Ground Truth

The dataset used in this project was curated by the NIH. It is comprised of 112,120 X-Ray images with disease labels from 30,805 unique patients. The disease labels for each image were created using Natural Language Processing (NLP) to process associated radiological reports. The estimated accuracy of the NLP labeling accuracy is estimated to be >90%.
This limitation on the NLP accuracy cannot be rectified, because the original radiology reports are not publically available to review and confirm labels. If these reports are available for human review, then the labels' accuracy will be reasonably higher.

6. FDA Validation Plan

Patient Population Description for FDA Validation Dataset:
The patient population should consist of Males and Females in a ratio of approximately 1:1, of ages 95 and younger, and are being screened for Pneumonia. The expected rate of Positive Pneumonia cases is approximately 20% of those being screened.
Imaging modality must be X-Rays (Code DX) of frontal positions (PA or AP) on Patient Chests.

Ground Truth Acquisition Methodology:
The purpose of this device is to assist Radiologist with their workflow. The optimal ground truth for this purpose would be the Radiologist's review of the X-Ray with medical history available to them, including lab reports. To bypass inaccuracies introduced by NLP, a digital form should be provided to Radiologists to mark diseases that they see with each X-Ray review. The checklist would included common lung diseases such as the 14 diseases in this dataset.

Algorithm Performance Standard:
Based on the provided paper by P.Rajpurkarpar, et al., the performance standard is F1 scores comparing radiologists and algorithms. F1 scores are the harmonic average of the precision and recall of the models.
Rajpurkarpar's CheXNet algorithm achieve an F1 score of 0.435, while Radiologists averaged an F1 score of 0.387. These would be the benchmarks to compare newly developed algorihms to.