Skip to content

This depository stores operation guidance for LC-MS/MS using the Thermo Scientific Vanquish UHPLC -- TSQ Altis systems and a custom data analysis toolkit.

License

Notifications You must be signed in to change notification settings

CreLox/LC-MS-MS

Folders and files

NameName
Last commit message
Last commit date

Latest commit

3364b6c · Aug 27, 2024

History

26 Commits
May 20, 2023
Mar 18, 2024
Aug 27, 2024
Mar 18, 2024

Repository files navigation

Column

The UHPLC column used for isolating steroids/secosteroids is Waters ACQUITY UPLC HSS T3 Column (100Å, 1.8 µm, 2.1 mm × 100 mm, Part# 186003539).

Flow rate

When setting up an LC-MS/MS run, make sure that TSQ Altis (Figure 2) is correctly informed of the LC flow rate (Figure 1).

image
Figure 1. LC flow rate in Xcalibur Instrument Setup (pump settings).


image
Figure 2. LC flow rate fed to TSQ Altis (ion source properties) in Xcalibur Instrument Setup (NOT in FreeStyle).

Vanquish pump

When an LC-MS/MS sequence finishes running, the vanquish pump should be stopped manually: go to "Direct Control", click "Take Control" at the top, and uncheck the "Flow" box under the "Vanquish Pump" tab.

image
Figure 3. Stopping the vanquish pump.

About

This depository stores operation guidance for LC-MS/MS using the Thermo Scientific Vanquish UHPLC -- TSQ Altis systems and a custom data analysis toolkit.

Resources

License

Stars

Watchers

Forks

Languages